Ring Test Evaluation of the Repeatability and Reproducibility of a Porcine Reproductive and Respiratory Syndrome Virus Oral Fluid Antibody Enzyme-Linked Immunosorbent Assay

Document Type

Article

Publication Date

10-2012

Abstract

The precision of a Porcine reproductive and respiratory syndrome virus (PRRSV) oral fluid antibody enzyme-linked immunosorbent assay (ELISA) was evaluated by calculating reliability coefficients for assay repeatability (within laboratory) and assay reproducibility (between laboratories). Randomly ordered oral fluid samples of known (n= 39) and unknown (n = 224) PRRSV antibody status were tested in 12 diagnostic laboratories. Each laboratory tested the samples twice, first using an antibody ELISA kit and reagents provided to them (phase 1) and then using an ELISA kit and reagents configured in their respective laboratory (phase 2). Repeatability (within laboratory) reliability coefficients calculated using results from samples of known PRRSV antibody status ranged from 0.724 to 0.997 in phase 1 and from 0.953 to 0.998 in phase 2. Reproducibility (between laboratories) reliability coefficients were calculated for 3 conditions: case 1—samples of unknown status (n = 224); case 2—samples of known status (n = 39), and case 3—all samples (n = 263). Among the 3 cases, reliability coefficients ranged from 0.937 to 0.964 in phase 1 and from 0.922 to 0.935 in phase 2. For case 3, it was estimated that 96.67% of the total variation in phase 1 and 93.21% in phase 2 could be attributed to the oral fluid samples themselves. Overall, the PRRSV oral fluid antibody ELISA was highly repeatable and reproducible. The current study supports the routine use of this test in laboratories providing diagnostic service to pig producers.

Publication Title

Journal of Veterinary Diagnostic Investigation

Volume

24

Issue

6

First Page

1057

Last Page

1063

DOI of Published Version

10.1177/1040638712457929

Publisher

Sage Publications

Rights

Copyright © 2012 Sage Publications

Share

COinS